• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 45; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 45; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 199721745
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/21/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during the a surgery on (b)(6) 2019, dr.(b)(6) became aware that the inner thread of screw was defective.The surgeon had to remove the screw and rod, because the locking screw could not screw in.A new screw has been inserted and a surgery delay of 30 min occured.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4).Udi: (b)(4).Visual examination of the screw revealed that the threads of the tulip head were torn off completely.Some pieces of the torn thread of the tulip head were stuck in between the threads of correction key.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn thread of the tulip head cannot be determined.Noted damage suggests that inadvertently cross threading of the correction key occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE SCREW 7.0 X 45
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8370962
MDR Text Key137170813
Report Number1526439-2019-51385
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034442105
UDI-Public(01)10705034442105
Combination Product (y/n)N
PMA/PMN Number
K142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number199721745
Device Catalogue Number199721745
Device Lot Number102633
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-