MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problem Use of Device Problem (1670)
Patient Problems Aortic Dissection (2491); No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2019
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, low oxygenation occurred.As per user facility, during the circulatory arrest, (aortic dissection circulatory arrest, hypothermia at 25°c, total duration 263.33mn, aortic clamping 166.29, assistance only 94.53 cerebroplegia 92.24) no oxygenation or decarboxylation problems, the blood gases were correct, flow rate at 1.5l for 20-30 minutes, then in front of a slightly lowered nirs (near infrared spectroscopy), increase the flow rate to 1.7l until the end of the circulatory arrest.The surgeon then advised to resume the flow, signing the end of the circulatory arrest, so the fio2 was set at 100% on the sechrist, increased the sweep and requested act which was greater than 400.The flow was resumed for over a few minutes.It was observed on the cdi that the parameters are panicking: the pao² decreases despite the fio² to 100% and the pco² rises but remains "reasonable", then the anesthesiologists put 100% fio² on the ventilator.In a few seconds: we move on to galloping hypercapnia still the pao2 decreases inversely with pco2.The sweeping on the sechrist was increase to 10 or 12 l.Cardioplegia was done, then the blood in all the tubing is black, pao² goes down to about 50mmhg on the cdi, pco² is 110mmhg.The parameters on the sechrist was kept at maximum fio² 100% and scan 12.Cec was changed with no problem on it.However, while they debubbled the 2nd cec it still has no improvements on the gases on the cdi.Results correlated by the blood gases that were done very closely as soon as they started to no longer properly oxygenate the patient.The product was changed out.It is unknown if there was a blood loss.It is unknown if there was a delay and if the procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 3331, 213, 67).Method code #1: 10 - testing of actual/suspected device.Method code #2: 3331 - analysis of production records.Results code: 213 - no device problem found.Conclusions code: 67 - no problem detected.The actual sample was received and was visually inspected upon receipt with no obvious anomalies such as break in the appearance.After having been rinsed and dried, the actual sample was tested for its gas transfer performance in accordance with the factory's shipping inspection protocol.No anomalies were revealed in the gas transfer performance of the actual sample, with the obtained values meeting the factory specifications.The investigation result verified that the actual sample, after having been rinsed, was the normal product with no issue in the gas transfer performance.From the available information, the possibilities can be inferred; however, it is difficult to determine the definitive root cause.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 27, 2019.Upon further investigation of the reported event, the following information is new and/or changed: g4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H6 (identification of evaluation codes 4114, 3221, 4315).Method code: 4114 - device not returned.Results code: 3221 - no findings available.Conclusions code: 4315 - cause not established.The actual sample was not returned for investigation and lot number was not provided.Multiple attempts were made to obtain additional information.Without further information, it is not possible to determine a definitive root cause.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.  upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a third follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8373453
MDR Text Key139187443
Report Number1124841-2019-00056
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Was the Report Sent to FDA? No
Date Manufacturer Received05/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SECHRIST
Patient Outcome(s) Required Intervention;
-
-