HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC Back to Search Results
Model Number P3700C000010
Device Problem Device Slipped (1584)
Patient Problem No Patient Involvement (2645)
Event Date 02/16/2019
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the brake detent mechanism needed to be replaced.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the brake detent mechanism to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the bed's brakes were not holding.The bed was located in the labor and delivery department at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key8373812
MDR Text Key138512942
Report Number1824206-2019-00082
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3700C000010
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-