Model Number 175810 |
Device Problems
Inflation Problem (1310); Difficult to Remove (1528)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the catheter was difficult to remove after the saline was withdrawn from the balloon.The catheter was removed by using a lubricant and manual manipulation.The patient voided after the removal without difficulty.Once removed, the nurse tried to fill the balloon again with saline, and noticed the balloon was only inflating on one side.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
|
|
Event Description
|
It was reported that the catheter was difficult to remove after the saline was withdrawn from the balloon.The catheter was removed by using a lubricant and manual manipulation.The patient voided after the removal without difficulty.Once removed, the nurse tried to fill the balloon again with saline, and noticed the balloon was only inflating on one side.
|
|