On (b)(6) 2019, the patient underwent a bronchoscopic lung volume reduction (blvr) procedure with zephyr endobronchial valve (ebv) placement.A zephyr 4.0-j endobronchial delivery catheter (edc) was used to size the airways and successfully deployed a zephyr valve.The zephyr 4.0-j edc was removed from the bronchoscope with the bronchoscope still in position in the patient.A second zephyr 4.0 valve was loaded into the zephyr 4.0-j edc, and the catheter was inserted into the bronchoscope which was articulated to reach lb1-2.While trying to advance further, resistance was met, and the catheter withdrawn.At that time, it was noticed that the housing on the zephyr 4.0-j edc was damaged but still attached, and one of the four sizing wings (also called diameter gauges) was detached.No device fragment had exited the bronchoscope and due to mucus, there were multiple suctions performed through the bronchoscope.The conclusion from the treating physician was that if the sizing wing detached in the bronchoscope it was removed during suction because there was no device fragment remaining in the patient.A new zephyr 4.0-j edc and zephyr 4.0 valve were used, and there were no further issues.
|