Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation therefore an analysis was not performed.X-ray images were provided and revealed that one of the rods there was separation of the lead screw from the magnet module from the rod on the right, which indicates a broken locking pin.The x-ray images of the left rod revealed lengthening and then loss of distraction in the rod on the left.Separation of the lead screw from the magnet module was not observed, therefore pin breakage could not be confirmed.Review of the device history record for the rod could not be performed due to insufficient information.
 
Event Description
It was reported that allegedly the patient's magec rods appeared to be damaged.The rods remain implanted, and proceeding to posterior spine fusion early summer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer Contact
karen jaffe
101 enterprise, suite 100
aliso viejo, CA 92656
9495446507
MDR Report Key8374645
MDR Text Key137294397
Report Number3006179046-2019-00022
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received02/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age15 YR
-
-