• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number MS1-5590S
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
A visual inspection of the returned devices revealed to be partially distracted with score marks on the distraction rods.Both rods could be lengthened by hand when pulling on the distraction rod, which confirms the pin failures.However, the repeated pistoning motion of the distraction rods, due to pin failures, may cause unintended wear of the o-rings.A review of the lot history record for the device revealed that the device met all of the required quality inspections and that the products were released within specifications.
 
Event Description
It was reported that allegedly the patient's magec rods were not getting anymore distraction after 4 (four) years of implantation.The physician elected to fuse the patient without incident.However, upon explanting the rods, appeared to be damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer Contact
karen jaffe
101 enterprise, suite 100
aliso viejo, CA 92656
9495446507
MDR Report Key8376147
MDR Text Key139197600
Report Number3006179046-2019-00025
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberMS1-5590S
Device Catalogue NumberPA0518
Device Lot NumberA140313-17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age12 YR
Patient Weight56
-
-