Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 00500105026, liner 26 mm i.D.For use with 50/51/52 mm o.D.Shells, 63999072.(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the customer was not able to insert the liner to the metal shell of multipolar.This surgery was finished with another shell.Attempts have been made and additional information on the reported event is unavailable.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Udi: (b)(4).Complaint sample was evaluated and the reported event was not confirmed.Visual inspection identified that the lock ring tab had moved and was outside the shell tab window in the lock ring groove.Gouge damage marks were noted to the lock ring tab.No other damages were observed.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|