• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZMR AIR DERMATOME HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZMR AIR DERMATOME HANDPIECE Back to Search Results
Catalog Number 00880100100
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 02/07/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation and investigation is in process.Once the investigation is complete, a supplemental medwatch will be filed.
 
Event Description
Repair request was reported because the screws that maintain the plates are missing.Event timing was before surgery.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
Result code- screws missing from the dermatome.The device history record for zimmer air dermatome serial number (b)(4) was reviewed noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The record review found no issues with the device and all verifications, inspections, and tests were successfully completed.Using the machine-repair reports and the repair sites folders on livelink as well as sap to query for all repairs on serial number (b)(4) prior to (b)(6) 2019, the device was noted to have not been previously repaired.The reported event was confirmed by the service technician who performed the evaluation and repair.On (b)(6) 2019, it was reported from (b)(6) that a dermatome was missing a width plate screw.The customer returned a zimmer air dermatome serial number (b)(4) for evaluation.Evaluation of the device on 13 march 2019 found that the device was out of calibration at the zero setting and that the width plate screws were missing.Repair of the device occurred on 30 april 2019 and involved replacing the missing screws and the needle bearing as well as recalibrating the device.The technician then tested and verified that the dermatome was functioning as intended, then returned the device to the customer without further incident.The device was tested, inspected, and repaired.While the service technician found that the width plate screws were missing from the device, it cannot be determined from the information provided as to what caused the screws to go missing.Therefore, a specific root cause of the reported event cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Event Description
No additional event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZMR AIR DERMATOME HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key8379743
MDR Text Key137462599
Report Number0001526350-2019-00143
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00880100100
Device Lot Number33163800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2019
Was the Report Sent to FDA? No
Date Manufacturer Received06/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-