The device was returned to olympus for evaluation.A visual inspection was performed on the returned device and found 90 percent of the ceramic beak from the distal end broken off and missing.The broken ceramic beak was not returned for evaluation.There is a small portion of the ceramic beak still intact at the inner portion of the sheath.Further inspection found severe dents on the outer tube at the distal end where the breakage occurred.Additionally, there were multiple dents and scratches noted throughout the whole outer tube.Based on the investigation findings, the broken/missing ceramic beak and dents on the outer tube are due to impact damage attributed to mishandling.The instruction manual for use states, "do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use.".
|
Olympus was informed during a therapeutic transurethral resection of bladder tumor (turbt) procedure, the plastic tip of the inner sheath broke off and fell into the patient¿s bladder.The device fragment was retrieved completely with a grasper.There was no unusual bleeding observed.It was reported that the doctor was resecting the tumor, when the breakage occurred.No x-ray was performed.No resistance was felt upon insertion of the device and no force was applied or required.The intended procedure was completed with a similar device.There was no patient injury reported.Additionally, the user facility reported that the sheath was inspected prior to use and no anomalies found.There were no issues noticed with the alignment and connection of the sheath when it was laid out in preparation for the procedure.The storage environment does not have heavy light exposure.The devices are stored in a room with no windows with normal inside lighting.
|