• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problem High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 11/21/2018
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Device 4 of 4.Reference mfr report # 1627487-2019-02524, reference mfr report # 3006705815-2019-00672, reference mfr report # 1627487-2019-02525.It was reported the patient experienced ineffective stimulation due to inability to increase amplitude.System diagnostics revealed high impedances on all contacts.Reprogramming initially was able to provide effective stimulation.Later the patient experienced ineffective stimulation post an cervical facet injection.In turn, the patient underwent surgical intervention wherein both leads were explanted and replaced.Therapy was restored post operatively.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Reference mfr report # 1627487-2019-02524; reference mfr report # 3006705815-2019-00672; reference mfr report # 1627487-2019-02525.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8380882
MDR Text Key137503679
Report Number1627487-2019-02526
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2016
Device Model Number3383
Device Lot Number4489115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/10/2019
Initial Date FDA Received03/01/2019
Supplement Dates Manufacturer Received07/01/2019
Supplement Dates FDA Received07/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3228, SCS LEAD; MODEL 3383, SCS EXTENSION; MODEL 3789, SCS IPG
Patient Outcome(s) Other;
Patient Age48 YR
Patient Weight50
-
-