Model Number 3383 |
Device Problem
High impedance (1291)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 02/12/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
Device 2 of 3: reference mfr report#1627487-2019-02997, reference mfr report#1627487-2019-03002.It was reported that the patient was unable to experience adequate stimulation.Lead diagnostics revealed high impedances on most of their contacts.Surgical intervention is pending to address the issue.
|
|
Manufacturer Narrative
|
The implant date was inadvertently missing in the initial report.The implant date is added to the additional report-1.
|
|
Event Description
|
Device 2 of 3: reference mfr report #1627487-2019-02997.Reference mfr report #1627487-2019-03002.Additional information received identified that surgical intervention was undertaken wherein the scs system was explanted and replaced.Reportedly the lead was fractured in multiple places, with one contact out of the lead completely.Therapy was restored post-operatively.
|
|