Model Number 3383
Device Problem High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/12/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Device 2 of 3: reference mfr report#1627487-2019-02997, reference mfr report#1627487-2019-03002.It was reported that the patient was unable to experience adequate stimulation.Lead diagnostics revealed high impedances on most of their contacts.Surgical intervention is pending to address the issue.
 
Manufacturer Narrative
The implant date was inadvertently missing in the initial report.The implant date is added to the additional report-1.
 
Event Description
Device 2 of 3: reference mfr report #1627487-2019-02997.Reference mfr report #1627487-2019-03002.Additional information received identified that surgical intervention was undertaken wherein the scs system was explanted and replaced.Reportedly the lead was fractured in multiple places, with one contact out of the lead completely.Therapy was restored post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8386850
MDR Text Key137708209
Report Number1627487-2019-03001
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734402323
UDI-Public05414734402323
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2016
Device Model Number3383
Device Lot Number4864024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2019
Initial Date FDA Received03/04/2019
Supplement Dates Manufacturer Received03/20/2019
Supplement Dates FDA Received03/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3219, SCS LEAD; MODEL 3383, SCS EXTENSION
Patient Outcome(s) Other;
Patient Weight61
-
-