• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. DVR 2.2MM DRILL BIT STERILE; TRAUMA, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. DVR 2.2MM DRILL BIT STERILE; TRAUMA, INSTRUMENT Back to Search Results
Catalog Number 212000022
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Patient Involvement (2645)
Event Date 02/06/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source - foreign - (b)(6).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Product not returned.
 
Event Description
It was reported that our incoming inspection member found debris in the sterile package.There was no patient involvement.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product identified that there is a piece of packaging debris inside the sterile package.Device history record was reviewed and no discrepancies were found.The product was likely non-conforming when it left zimmer biomet control.The root cause of the reported event can be attributed to the operator not following instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DVR 2.2MM DRILL BIT STERILE
Type of Device
TRAUMA, INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8392050
MDR Text Key137887004
Report Number0001825034-2019-00971
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number212000022
Device Lot Number949190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/06/2019
Initial Date FDA Received03/05/2019
Supplement Dates Manufacturer Received08/29/2019
Supplement Dates FDA Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-