Model Number FF494R |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).When additional information is received a follow up report will be submitted.
|
|
Event Description
|
It was reported by the healthcare professional "the surgeon reports that the closing system applicator clamp attachment in placing the craneofix implant does not make the correct tightening, leaving the implant without sufficient force." there was a 15 minute delay in surgery when placing the third implant.
|
|
Manufacturer Narrative
|
Investigation: used test and analysis equipment: keyence vhx-5000 digital microscope.Panasonic dmc tz8 digital camera.Mecmesin afg 500n force measuring device.For investigation we sent the instrument to our technical service department.There they tested the clamping function and the pulling force of the device.The clamping function worked well.All measured values are within the specification between 280 newtons and 350 newtons.Conclusion and root cause: the root cause of the problem is most probably usage related.Rationale: we tested the function and the force values of the instrument and found it according to the specification.Without further knowledge about the circumstances, we assume a handling error (not correctly attached to the pin) or a bent/defective pin.No capa is necessary.
|
|
Search Alerts/Recalls
|