Device 3 of 4: reference mfr report #3006705815-2019-00472.Reference mfr report #3006705815-2019-00473.Reference mfr report #1627487-2019-02830.It was reported that the patient could not place the ipg in the surgery mode.During reprogramming the leads displayed high impedances.Also, x rays showed the leads were potentially fractured.As a result, surgical intervention was undertaken wherein the leads and the anchors were explanted and replaced.Therapy was restored post-operatively.The physician had cut the anchors during the original implant and therefore the anchors are made reportable.
|