• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number 165816
Device Problems Material Invagination (1336); Material Deformation (2976)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported that the balloon of the silicone foley catheter mushroomed.
 
Manufacturer Narrative
The reported event was unconfirmed since the reported failure could not be reproduced.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley with an attached meter bag, sample port connector, and inlet tubing.The visual inspection of the catheter surface noted no mushroom on the balloon.The catheter balloon was inflated with 10ml of methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) and balloon concentricity was observed to be 60:40.The balloon rested for 30 minutes without leaks and passively deflated without issue, returning 10ml of solution with no cuffing noted.The active length of the catheter balloon was measured (0.7365") and found to be within specification (0.6" - 0.9").The catheter was confirmed to be 16 fr.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 30cc balloon: use 35ml sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.".
 
Event Description
It was reported that the balloon of the silicone foley catheter mushroomed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Type of Device
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8400633
MDR Text Key138360422
Report Number1018233-2019-01203
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741029752
UDI-Public(01)00801741029752
Combination Product (y/n)N
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 05/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/23/2023
Device Model Number165816
Device Catalogue Number165816
Device Lot NumberNGCV0822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2019
Initial Date Manufacturer Received 02/14/2019
Initial Date FDA Received03/07/2019
Supplement Dates Manufacturer Received05/10/2019
Supplement Dates FDA Received05/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-