Model Number 4296 |
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: 6935 lead.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that oversensing was seen on the ventricular channel during manual impedance measurement testing.Electrograms show noise on all channels.This led to pacing inhibition and a short asymptomatic pause.The physician is also wondering if the chroniccapped redundant right ventricular lead could be a factor in the oversensing.Both the right ventricular and left ventricular leads remain in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|