• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4570S
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2017
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation therefore an analysis was not performed.An inspection of the x-ray images provided revealed that one of the locking pins of the implanted rods is damaged.A review of the dhr determined that both of the rods had the old locking pin which has been improved.A review of the lot history record for the device revealed that the device met all of the required quality inspections and that the products were released within specifications.
 
Event Description
It was reported that allegedly one of the patient's magec rods is not lengthening.The physician has no plans to remove the magec rods at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer Contact
claudia
101 enterprise, suite 100
aliso viejo, CA 92656
9496528012
MDR Report Key8414680
MDR Text Key138806314
Report Number3006179046-2019-00017
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMS1-4570S
Device Catalogue NumberPA0561
Device Lot NumberA141210-03-00-01U, A150127-01-
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/11/2017
Initial Date FDA Received03/12/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-