Catalog Number 725.52
Device Problems Failure to Cut (2587); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
A sample is available that has not yet been received at manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported that two ophthalmic scissors cut poorly and failed to open or close during vitrectomy surgery.An alternate scissors was obtained in order to complete the procedure.Patient impact information is unknown.
 
Manufacturer Narrative
Four unopened samples were received in an opened box.The samples were functionally inspected with the aid of a photomicroscope with various magnifications.Three of the instruments meet the specifications.One sample does not smoothly open and close and, due to this, the cutting is poor.The customer¿s complaint was confirmed.It resulted that the instrument does not smoothly open and close so the cutting is poor.The instrument has maybe been incorrectly adjusted.This is a single case so no other action will be taken.The device history record for the suspected lot was reviewed.No abnormalities that could have contributed to this event were found and the product was released according to acceptance criteria.A 100% final inspection is performed for this product.The insufficient adjustment could lead to a lack of functionality and has not been noticed during manufacturing.As it is a single event, no further actions will be defined.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER ADVANCED DSP TIP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
MDR Report Key8417030
MDR Text Key138952802
Report Number3003398873-2019-00013
Device Sequence Number1
Product Code HNF
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Catalogue Number725.52
Device Lot NumberF160680
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Date Manufacturer Received03/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-