Catalog Number ER320
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r93u8f.Device analysis: the er320 device was returned for analysis and upon inspection the jaws were found to be in a yielded condition.In an attempt to replicate the reported incident the device was functionally evaluated.Upon firing of the device, the remaining 2 clips were ejected due to the condition of the jaws; finally the instrument locked out as intended.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.Possible causes for the found condition of the yielded jaws may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.It is known from the history of the device that the condition of the jaws may lead dropping/ejected clips.A manufacturing record evaluation was performed for the finished device lot number r40p0w, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number r93u8f, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic unknown procedure, the clip was fed into the jaws sideways during use.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8417961
MDR Text Key138795411
Report Number3005075853-2019-17232
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Catalogue NumberER320
Device Lot NumberR40P0W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2019
Initial Date Manufacturer Received 02/19/2019
Initial Date FDA Received03/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2018
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-