Event summary: patient data files showed at least eighteen applications were performed with catheter 2af284/ 66896-115 without any issue on the date of the event.In conclusion, a clinical issue occurred during the case.There is no indication of relation of adverse event to the performance of the cryo device.Data files cannot assess the steerability issue.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryoablation procedure, st elevation was observed.Nitrogen and dopamine were injected intravenously with resolve, and an angiography was performed.No stenosis or air was observed.The case was completed with cryo.It was unknown whether the event led to or extended patient hospitalization.No further patient complications have been reported as a result of this event.
|