• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100014514
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/28/2019
Event Type  Injury  
Event Description
During an ablation procedure, the amplifier would not function beyond the self test and the procedure was cancelled.Troubleshooting with the power cord, power outlet and air filter were performed, with no resolution.
 
Event Description
There were no adverse patient consequences due to the cancellation.
 
Manufacturer Narrative
Additional information: no product was returned for investigation. the device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott, the reported cancellation could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key8422666
MDR Text Key138852627
Report Number2184149-2019-00026
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number100014514
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2019
Initial Date FDA Received03/14/2019
Supplement Dates Manufacturer Received03/18/2019
Supplement Dates FDA Received03/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-