HOLOGIC, INC SELENIA MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SEL-00001
Device Problem Unintended System Motion (1430)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/28/2019
Event Type  malfunction  
Event Description
It was reported that uncontrolled gantry rotation had occurred.No injury reported.A field engineer was dispatched to the site.It was determined that the rotational switch was stuck at times.The rotation switch was readjusted and once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key8429479
MDR Text Key139098640
Report Number1220984-2019-00023
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSEL-00001
Device Catalogue NumberSEL-00001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/28/2019
Initial Date FDA Received03/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-