• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-062
Device Problem Mechanical Problem (1384)
Patient Problems Iatrogenic Source (2498); Device Embedded In Tissue or Plaque (3165)
Event Date 02/19/2019
Event Type  Injury  
Manufacturer Narrative
Device remains within patient.Therefore, no device evaluation will be performed.Instructions for use for this device state: warning: excessive applied traction forces may impact the lld¿s ability to disengage from a lead.
 
Event Description
A philips representative reported that a cardiac lead management procedure commenced to extract 3 non-functional/redundant leads.The physician prepped 2 of the leads for extraction (1 right atrial and 1 right ventricular) with a spectranetics lead locking device (lld) ez 518-062.The physician decided to abandon the extraction of the rv and ra leads after several attempts with no progress at an adhesion sight in the superior vena cava (svc).He attempted to unlock the lld ez, but was unable to and the lld was cut, capped within the respective lead, and left within the patient.The physician did not complete the lead extraction for any of the leads.However, he did manage to get access and complete the re-implant of a biv device.This mdr is being filed to capture the lld cut and capped within the ra lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
catherine eaton
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key8430446
MDR Text Key139164205
Report Number1721279-2019-00047
Device Sequence Number1
Product Code DRB
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,02/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2019
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP17L13A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received02/19/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
Patient Weight82
-
-