(b)(4).Only event year known: 2019.Batch # r9538g.Device analysis: the analysis results found that the el5ml device was returned with no damage in the external components and a plastic bag with a scissored clip inside.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 8 conforming clips.Upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch number r9538g, and no non-conformances related to the reported complaint condition were identified.
|