MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number 37603 |
Device Problems
Electromagnetic Interference (1194); No Device Output (1435); Device Difficult to Program or Calibrate (1496); Data Problem (3196)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Date 03/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins) for movement disorders.It was reported the patient had a post-op mri check, and the technician (tech) used an 8840 to turn the left ins to 0v and off.The tech could not turn the ins back on after the procedure, and when the programming was checked no electrodes were active.They saw 100 usec as the pulse width but the setting on record was 70 usec.The 8840 would not allow them to program 2-1- with c+, and there was only a greyed-out button.The patient had a return of symptoms due to the ins being off for the mri.The caller confirmed the patient was in current mode when they should have been in voltage mode.It was confirmed no interleaving was occurring.The caller was able to switch the patient back to voltage mode, turned the system on and the patient's symptoms were now under control.It was stated the caller was ultimately unsure if the programming changes were done by the mri tech or an ins issue.No further complications were reported or anticipated with the event.
|
|