• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37603
Device Problems Electromagnetic Interference (1194); No Device Output (1435); Device Difficult to Program or Calibrate (1496); Data Problem (3196)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 03/15/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins) for movement disorders.It was reported the patient had a post-op mri check, and the technician (tech) used an 8840 to turn the left ins to 0v and off.The tech could not turn the ins back on after the procedure, and when the programming was checked no electrodes were active.They saw 100 usec as the pulse width but the setting on record was 70 usec.The 8840 would not allow them to program 2-1- with c+, and there was only a greyed-out button.The patient had a return of symptoms due to the ins being off for the mri.The caller confirmed the patient was in current mode when they should have been in voltage mode.It was confirmed no interleaving was occurring.The caller was able to switch the patient back to voltage mode, turned the system on and the patient's symptoms were now under control.It was stated the caller was ultimately unsure if the programming changes were done by the mri tech or an ins issue.No further complications were reported or anticipated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8432582
MDR Text Key139204862
Report Number3004209178-2019-05568
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169864214
UDI-Public00643169864214
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2019
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2019
Date Device Manufactured03/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
-
-