Model Number H1-M |
Device Problem
Fracture (1260)
|
Patient Problem
Patient Problem/Medical Problem (2688)
|
Event Date 03/15/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the physician used a h1-m hawkone device with a non-medtronic 6frx55 sheath and 0.014x6.0 spider fx embolic protection device during treatment of a 60mm plaque 70% stenotic lesion in the patient¿s distal left superficial femoral artery (sfa) of diameter 5mm.Little tortuosity and little calcification were reported.Ifu was followed and the device was prepped without issue.The vessel was not pre-dilated.It was reported that the nosecone became detached from the body of the catheter but stayed attached to the filter wire.The nose cone and tip were snared successfully.A drug coated balloon was used to complete the procedure.No patient injury was reported.
|
|
Manufacturer Narrative
|
Additional information: it was reported that the tip detached upon retrieval just prior to the sheath in the vessel, not in the treated lesion.The hinge pin remained attached.The snare used was a non-medtronic device.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|