• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Ischemia (1942); Myocardial Infarction (1969); Thrombosis (2100); Prolapse (2475)
Event Date 02/25/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Date of implant has been estimated.The stent remains in the patient.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the patient presented emergently with a st segment elevation myocardial infarction (stemi).An absorb scaffold had been implanted 3-4 years prior; angiography revealed plaque rupture, some thrombus, and timi 2 flow in the proximal ramus artery.The lesion was direct stented with a xience xpedition device for treatment.There was no reported adverse patient sequela.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).Unique device identifier (udi): in the absence of reported part number, udi cannot be calculated.The stent remains in the vessel; however, there was no reported device malfunction and the product was not returned.A review of the lot history record and complaint history of the reported lot could not be conducted because the part and lot numbers were not provided.The reported patient effects of ischemia, myocardial infarction and thrombosis, are listed in the absorb gt1 instructions for use as known adverse events associated with the use of a coronary scaffold in native coronary arteries.Based on the case information, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8436803
MDR Text Key139336344
Report Number2024168-2019-02156
Device Sequence Number1
Product Code PNY
Combination Product (y/n)N
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/25/2019
Initial Date FDA Received03/20/2019
Supplement Dates Manufacturer Received03/22/2019
Supplement Dates FDA Received03/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability;
-
-