Device 3 of 3: reference mfr.Report#: 1627487-2019-03722; reference mfr.Report#: 1627487-2019-03723.It was reported one of the patient's lead tails was completely out of one of the extensions and the other lead was almost out of the extension (confirmed via x-rays).In turn, the patient underwent surgical intervention on (b)(6) 2019 wherein both extensions were revised.However, system diagnostic revealed invalid and high impedances.As a result, both extensions were explanted and replaced which resolved the issue.Therapy was restored post operatively.
|