• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 4L RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 4L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RW40C
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/04/2019
Event Type  Injury  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that 45 minutes while on cardiopulmonary bypass, the po2 dropped and the pco2 increased.These were not able to be corrected with fio2 or sweep gas.The user facility spliced another oxygenator into recirc line to resolve the issue.No consequences or impact to the patient.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 21, 2019.Upon further investigation of the reported event, the following information is new and/or changed: (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: (identification of evaluation codes 10, 11, 3331, 213, 67).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 213 - no device problem found.Conclusions code: 67 - no problem detected.The device was visually inspected upon receipt and it was found that the purge line tube and the sampling line tube had been cut off.The investigation result verified that the actual sample, after having been rinsed, was the normal product with no issue in the gas transfer performance, it is possible that the actual sample was used for the patient with large body surface area and the performance the fx15 size can serve was insufficient for the patient.From the available information and with no detailed information about the events leading up to the complaint or the absence of the involved perfusion record, the cause of the complaint cannot be determined.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15RWC W/ 4L RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8441434
MDR Text Key139537620
Report Number1124841-2019-00078
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450790
UDI-Public(01)00699753450790
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 06/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model Number3CX*FX15RW40C
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/21/2019
Supplement Dates Manufacturer Received04/17/2019
06/12/2019
Supplement Dates FDA Received05/02/2019
06/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-