• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. AZUR DETACHABLE 35 HYDROCOIL; EMBOLIZATION COIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR DETACHABLE 35 HYDROCOIL; EMBOLIZATION COIL Back to Search Results
Model Number MV-AZ50620HD
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 02/18/2019
Event Type  malfunction  
Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was discarded at the user facility; therefore, a product analysis could not be performed.The root cause is unknown.
 
Event Description
It was reported that during preparation, the embolization coil detached when it was soaked in a heated 60°c saline solution.The device was not used in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR DETACHABLE 35 HYDROCOIL
Type of Device
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
Manufacturer Contact
debby callahan
35 enterprise drive
aliso viejo, CA 92656
MDR Report Key8441797
MDR Text Key139551690
Report Number2032493-2019-00062
Device Sequence Number1
Product Code KRD
UDI-Device Identifier04987892097591
UDI-Public(01)04987892097591(11)180504(17)230430(10)18050451V
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K093002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberMV-AZ50620HD
Device Lot Number18050451V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2019
Initial Date FDA Received03/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-