Model Number 102
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2018
Event Type  malfunction  
Event Description
It was reported that the patient's settings were inadvertently "zeroed out" at a patient's previous appointment in december.The session report at the patient's last appointment in july, everything was fine.Programming history was reviewed and it was confirmed that the patient's normal mode output current was 0 ma upon interrogation in december; however, data from the appointment at july was unavailable.It is known that because 102 generators temporarily changes the programmed settings to run system diagnostics test, interrupted diagnostics may lead to unintended settings changes.In labeling, it indicates that final interrogations should be performed at the end of every appointment to confirm correct settings.It is suspected that an interrupted test occurred at the previous appointment and the physician did not confirm the settings with a final interrogation, but this has not yet be confirmed.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 102
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8445336
MDR Text Key139608427
Report Number1644487-2019-00578
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/18/2016
Device Model Number102
Device Lot Number202960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 02/27/2019
Initial Date FDA Received03/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/13/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
-
-