MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR VUE WIRELESS FOOTSWITCH; FOOT-SWITCH, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR VUE WIRELESS FOOTSWITCH; FOOT-SWITCH, ELECTRICAL Back to Search Results
Catalog Number 227214
Device Problem Intermittent Continuity (1121)
Patient Problem Not Applicable (3189)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter phone number: (b)(6).Udi: (b)(4).
 
Event Description
It was reported by the affiliate via email that during shoulder arthroscopy procedure the coagulation pedal on the vapr vue wireless footswitch was working intermittently.The device was not able to sustain function when pressed.The procedure was completed by changing the pedal with no patient harm, but a 5 minute time delay was reported.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).Investigation summary: the complaint device is not being returned, it was retained by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device (lot : 1611221), and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR VUE WIRELESS FOOTSWITCH
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8448610
MDR Text Key139725449
Report Number1221934-2019-56674
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705009657
UDI-Public10886705009657
Combination Product (y/n)N
PMA/PMN Number
K100638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227214
Device Lot Number1611221
Was Device Available for Evaluation? No
Date Manufacturer Received04/24/2019
Patient Sequence Number1
-
-