• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number 544995
Device Problems Leak/Splash (1354); Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/04/2019
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the jaws did not open and close properly because of misalignment during use.It was replaced with a new kit.
 
Manufacturer Narrative
Qn# (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.Kenosha, wi facility as part of a 50 pc.Lot in february of 2018.The returned instrument as received was evaluated and found that the jaws are not misaligned as stated in complaint and this instrument is able to pick-up, retain, close and release multiple clips both with and without the use of silastic test tubing.We are unable to validate this complaint since we the jaws are aligned , and we are unable to locate any non-conforming features on the returned sample part.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.
 
Event Description
It was reported that the jaws did not open and close properly because of misalignment during use.It was replaced with a new kit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL L 10MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
MDR Report Key8452974
MDR Text Key139889505
Report Number3011137372-2019-00095
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544995
Device Lot Number06D1750217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2019
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/26/2019
Supplement Dates Manufacturer Received04/15/2019
Supplement Dates FDA Received04/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-