We've been asked about the cleaning procedure and how can we be sure that the inside of the moving part has been properly cleaned as there is no apparent way to dissemble.In the first picture the moving part is retracted, in the second its not.The component normally sits as per the second picture and this is how it would be processed for cleaning.Even if it was manually done, there's still the inside of the sleeve that doesn't get cleaned.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: in response to the questions asked about cleaning the quickset ace grater handle, our sterility assurance team have stated that if the instructions for use ( ifu) are followed then there will not be a problem cleaning the instrument.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|