• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problem High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Event Description
Device 1 of 2; reference mfr report # 1627487-2019-03864.It was reported that patient had high impedances on contacts 2 and 8.In turn, surgical intervention was undertaken on (b)(6) 2019 wherein the extension was explanted and replaced, and an additional lead was added at t5 to improve coverage.The ipg was also electively replaced during the procedure.Effective therapy was restored postoperatively.
 
Event Description
Device 1 of 2: reference mfr report # 1627487-2019-03864.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8454231
MDR Text Key139952335
Report Number1627487-2019-03863
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017918
UDI-Public05415067017918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number3383
Device Lot Number5004244
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2019
Initial Date FDA Received03/26/2019
Supplement Dates Manufacturer Received04/16/2019
Supplement Dates FDA Received04/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight104
-
-