Image review: image 1 and 2 shows the filter observe outside of the distal section of the green dual ended delivery catheter.This likely occurred after the spiderfx device was removed out of the patient.The catheter was observed to be occluded with red biological material.The distal tip wire was present, and the coil shows no evidence of uncoiling.Filter assembly appears to be intact.Cine 1 shows the filter is fully deployed in the vessel.The proximal radiopaque marker band and the distal radiopaque marker band are visible.There is evidence of contrast in the targeted vessel.Base on the cine quality, we are unable to confirm a tip detachment and vessel dissection.However, based on image 1 and 2 received, the distal tip wire is present, and the coil shows no evidence of uncoiling.Filter assembly appears to be intact.If information is provided in the future, a supplemental report will be issued.
|