Catalog Number 111140
Device Problems Crack (1135); Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/27/2019
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The tip of the slaphammer broke off at the screw threads.
 
Manufacturer Narrative
An event regarding crack/fracture involving a mako slaphammer was reported.The event was confirmed by photograph.Method & results product evaluation and results: the reported device was not returned however photographs were provided for review.The photographs note the following: the threaded tip has broken off the end of the slaphammer at the based of the threaded section.There is nothing else remarkable to note.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as the primary operative report and device return are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
The tip of the slaphammer broke off at the screw threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL SLAPHAMMER
Type of Device
KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8457256
MDR Text Key140155013
Report Number3005985723-2019-00264
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486002732
UDI-Public00848486002732
Combination Product (y/n)N
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number111140
Device Lot Number06111015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2019
Initial Date Manufacturer Received 02/27/2019
Initial Date FDA Received03/27/2019
Supplement Dates Manufacturer Received10/15/2019
Supplement Dates FDA Received11/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-