• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CHEST COMPRESSION ASSEMBLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CHEST COMPRESSION ASSEMBLY Back to Search Results
Model Number 8700-0701-01
Device Problems Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/03/2019
Event Type  malfunction  
Manufacturer Narrative
The autopulse lifeband in complaint was disposed by the customer and it will not be returned for investigation.Therefore, a physical investigation will not be performed.A supplemental report will be filed if and when the lifeband is returned and investigation has been completed.
 
Event Description
The autopulse platform (sn (b)(4)) was used on a 80 year old female patient weighing 350 lbs.Approximately 10 mins.During compression the lifeband started to tear and twist and the autopulse stopped compression and autopulse displayed "realign patient".The crew performed troubleshooting steps by re-aligning the patient multiple times, however the platform stopped each time after performing several compressions.The use of the autopulse was discontinued and manual cpr was performed.No patient consequence.Additional information: per reporter the lifeband was disposed.See mfr 3010617000-2019-00296 for the issue with autopulse platform (sn (b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LIFEBAND
Type of Device
CHEST COMPRESSION ASSEMBLY
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave,
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave,
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key8457843
MDR Text Key140171755
Report Number3010617000-2019-00297
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001571
UDI-Public00849111001571
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0701-01
Device Catalogue Number8700-0701-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient Weight159
-
-