DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1006-9305-000
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The reported event was addressed by ge healthcare technical support by troubleshooting with the customer biomedical engineer.The reported event cannot be corrected remotely and the hospital has not called back for service.No further resolution information available.
 
Event Description
This report summarizes 1 malfunction event.A review of the event indicated that model 1006-9305-000 anesthesia gas machine had improper flow from a reported gas proportioning system not operating as expected and causing the potential for a hypoxic mix of gas delivery.This report was from a single source.This event did not involve a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7900
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key8471267
MDR Text Key145394274
Report Number2112667-2019-00029
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Type of Report Initial
Report Date 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number1006-9305-000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-