• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number C-VH-4000
Device Problems Intermittent Continuity (1121); Intermittent Loss of Power (4016)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A lot history record review was completed for the reported product lot number.There were no ncmr¿s for the reported lot number.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2.While performing a radial artery harvest ,about half way through burning branches they started to have intermittent power when pulling back the toggle.They did not feel it was a time out feature creating the power loss.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2.While performing a radial artery harvest, about half way through burning branches they started to have intermittent power when pulling back the toggle.They did not feel it was a time out feature creating the power loss.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Manufacturer Narrative
Internal complaint number: (b)(4).Autonumber: (b)(4).Specific actions for the analyzed failure mode are being maintained and documented under maquet's failure investigation report (fir) system.The device was returned to the factory for evaluation.Signs of clinical use was observed.There was evidence of blood detected.A visual inspection was conducted.Small amounts of tissue and charred blood were observed on the jaws, and heater wire.The heater wire was observed to be flexed away at the center of the hot jaw.The wire remained attached at the base and tip of the hot jaw.No other failures were observed.The device was evaluated for electrical function.A pre-cautery test was performed per the instruction for use with a reference cable, adapter and reference power supply at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.Based on the results of the evaluation, the reported failure ¿intermittent continuity¿ is not confirmed, however the analyzed failure "material twisted/bent; wire" is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
MDR Report Key8472950
MDR Text Key140623642
Report Number2242352-2019-00370
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2021
Device Catalogue NumberC-VH-4000
Device Lot Number25143949
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Date Manufacturer Received06/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-