The initial reporter stated that they noticed an issue with one reagent kit of the elecsys vitamin d total gen.2 assay on the cobas 8000 e 602 module.An unspecified number of patient samples resulted with values of > 250 nmol/l and when repeated on another analyzer, results were within the technical limit of the assay.The customer provided data for a total of 36 patient samples and of these, one had an erroneous vitamin d result.The erroneous result was not reported outside of the laboratory.The sample initially resulted with a value of 88.89 nmol/l when measured with the problematic kit.When tested on the other analyzer, the vitamin d result was 48.98 nmol/l.No adverse events were alleged to have occurred with the patient.The e 602 analyzer serial number is (b)(4).
|