Model Number 9-ASD-034
Device Problem Migration (4003)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2019
Event Type  Injury  
Event Description
On (b)(6) 2018, a 34mm amplatzer septal occluder was implanted.Three months following the implant, the patient presented with chest pain.A toe was performed and revealed the device had moved posteriorly but was not completely dislodged.The device was explanted and a surgical patch was placed.The patient remained hemodynamically stable throughout the procedure and was stable post-operatively.
 
Manufacturer Narrative
An event of a dislodged device was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8480062
MDR Text Key140830160
Report Number2135147-2019-00096
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010243
UDI-Public00811806010243
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model Number9-ASD-034
Device Catalogue Number9-ASD-034
Device Lot Number6281436
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2019
Initial Date FDA Received04/03/2019
Supplement Dates Manufacturer Received04/11/2019
Supplement Dates FDA Received04/19/2019
Date Device Manufactured01/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age23 YR
-
-