• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 METAGLENE POSITIONER PLATE; EXTREMITY INSTRUMENTS : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 METAGLENE POSITIONER PLATE; EXTREMITY INSTRUMENTS : ALIGNMENT DEVICES Back to Search Results
Catalog Number 230787003
Device Problems Break (1069); Material Deformation (2976); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during pinning, the metaglene pin broke off and was lodged in the baseplate.Pieces were retrieved and the case was not delayed.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned device identified the tip of the guide pin broke off and was jammed/seized in the device.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Added: d10.Corrected: h3.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METAGLENE POSITIONER PLATE
Type of Device
EXTREMITY INSTRUMENTS : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key8483045
MDR Text Key140916323
Report Number1818910-2019-89752
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295116448
UDI-Public10603295116448
Combination Product (y/n)N
PMA/PMN Number
K170748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number230787003
Device Lot Number5219831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2019
Initial Date Manufacturer Received 03/28/2019
Initial Date FDA Received04/04/2019
Supplement Dates Manufacturer Received04/08/2019
05/21/2019
Supplement Dates FDA Received05/03/2019
05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age77 YR
-
-