Model Number M0062201090 |
Device Problem
Burst Container or Vessel (1074)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an occluder ballon catheter was used in a procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the balloon burst.
|
|
Event Description
|
It was reported to boston scientific corporation that an occluder balloon catheter was used in a procedure performed on (b)(6), 2019.According to the complainant, during the procedure, the balloon burst.
|
|
Manufacturer Narrative
|
Problem code 1074 captures the reportable issue of balloon burst.Investigation results: a visual examination of the returned complaint device confirmed that the balloon burst.It was noticed that the proximal and distal bonds are continuous.Functional analysis could not be performed due to the returned condition of the balloon.There were no other issues noted with the device.This failure is likely due to factors and conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, th most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
|
Search Alerts/Recalls
|