Age at time of event: 18 years or older.Device evaluated by manufacturer: the device was returned for analysis.Returned product consisted of a renegade hi-flo micro-catheter stuck inside a terumo guide catheter.The device was analyzed as received.The device was stuck inside of a 4fr terumo guide catheter.The renegade device showed a fracture located 36.4cm from the hub.There was extreme tensile force applied to the device due to the stretching that was noticed.There were multiple kinks throughout the returned devices.The devices were soaked in an 37 degree celsius bath for a period of 72hrs to try and loosen the devices up to pull the renegade from the guide catheter; however, the devices could not be separated.A.014 fathom guidewire was attempted to be inserted into the hub end of the device; however, the guidewire would not enter due to contrast being hardened in the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that friction occurred.A 140cm/115cm renegade hi-flo fathom system was selected for use.During procedure, a non bsc catheter was used to perform celiac trunk angiography for mapping hepatic vascular anatomy.Subsequently, hi-flo microcatheter system was introduced through celiac trunk to proper hepatic artery; however, friction was very large.To find where the problem was, the wire was retracted and the hiflo catheter was re-introduced but the wire could not be introduced again into the catheter.Consequently, it was speculated that there was something wrong with the hub.The procedure was completed with different device.No patient complications were reported and the patient is stable.However, device analysis revealed device fracture.It was further reported that the fracture may have been caused by pulling or pushing the microcatheter with force.
|