• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS-320-K
Device Problem Failure to Interrogate (1332)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Exhibited a reset after implant subsequent to detection being enabled for the first time.The root cause of this failure has been traced to a random manufacturing defect in the logic circuit.
 
Event Description
The rns neurostimulator was placed without complications on (b)(6) 2019.However while attempting to interrogate the device in the recovery room the device displayed an alert message preventing further programming.A firmware update was applied to the device, which allowed the device to be interrogated and programmed.However the alert message reoccurred.A review of the device data determined that the rns neurostimulator underwent an unrecoverable error and would need to be replaced.The patient was in stable condition throughout the troubleshooting.The patient underwent a device replacement procedure without complications on (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8498194
MDR Text Key141492588
Report Number3004426659-2019-00009
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005267
UDI-Public010085554700526717191121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2019
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberRNS-320-K
Device Catalogue Number1007694
Device Lot Number27147-1-1-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age17 YR
-
-