ZOLL CIRCULATION INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Display Difficult to Read (1181)
Patient Problem No Patient Involvement (2645)
Event Date 03/19/2019
Event Type  malfunction  
Manufacturer Narrative
The reported event of "the user observed broken lcd display on the autopulse platform (sn (b)(4)) and the display was not readable" was confirmed during visual inspection.The most possible root cause for the reported issue was due to user mishandling.Upon visual inspection, observed broken and not readable lcd display and scratched user interface (ui) membrane.Unable to perform functional testing due to user advisory (ua) 45 (not at "home" position after power-on/restart) error message displayed upon powering on the device.The encoder drive shaft was not within the normally acceptable range when powered on.The drive shaft was returned to home position using the administrative menu.Upon customer approval, the defective parts will be replaced and the autopulse platform will be further tested to full specification.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with serial number (b)(4).
 
Event Description
During training, the user observed broken lcd display on the autopulse platform (sn (b)(4)) and the display was not readable.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave
san jose, CA 95131
4085411030
MDR Report Key8506335
MDR Text Key141740351
Report Number3010617000-2019-00276
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000536
UDI-Public00849111000536
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-