ABBOTT QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3367-40QC
Device Problem Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/21/2019
Event Type  malfunction  
Event Description
During a lead revision procedure, blood was noted in the header of the device.The device was successfully replaced, and the patient was stable with no consequences.Related manufacturer reference number: 2938836-2019-02304.
 
Manufacturer Narrative
Visual examination of the device found body fluid was present underneath the df4 septum and around the df4 connector block.The reported field event of blood in the header was confirmed, and was consistent with the revision procedure and is not considered a device failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8508369
MDR Text Key141732702
Report Number2938836-2019-02303
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberCD3367-40QC
Device Lot NumberP000070543
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/31/2019
Initial Date FDA Received04/12/2019
Supplement Dates Manufacturer Received09/28/2019
Supplement Dates FDA Received10/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION
-
-