Age at time of event: 18 years or older.Device evaluated by manufacturer: the device was returned for analysis.Returned product consisted of a renegade stc-18 micro-catheter.The device showed damage at the distal end from the tip proximal 14cm.The shaft looked flattened in some areas as well as some delaminating of the outside shaft.The device was very sticky and stuck together at the distal end.The shaft showed a kink located at 62.7cm from the hub.A.014 guidewire was attempted to be put through the distal tip; however, the damage would not allow the wire to transcend through the device.Inspection of the remainder of the device, except for the damage observed revealed no other damage or irregularities.
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that the fathom wire would pass through the catheter.A 130/20 renegade stc 18 was selected for use.During the procedure, it was noted that the fathom wire would not pass thorough tapered tip of the renegade catheter.Subsequently, the device was removed through the introducer sheath and no damage was noted on the fathom wire.The procedure was completed with another of the same device.No patient complications were reported and the patient was fine.However, device analysis revealed delamination of the outer shaft.
|